    Cardiac Myoglobin Uncut Sheet Rapid Test ISO13485 Rtk Blood Test

    Cardiac Myoglobin Rapid One Step Test with CE certificate uncut sheet

    Product Name

    Cardiac Myoglobin One Step Test

    FormatsStrip(3mm)Device(4mm) uncut sheet
    Place of OrigenChina
    SpecimenBlood
    Read Time15 minutes
    Shelf life2 years
    Packageuncut sheet
    Storage2℃-30℃

    For professional in vitro diagnostic use only.


    INTENDED USE

    The Myoglobin Rapid Test Device (Whole Blood/Serum/Plasma) is a rapid chromatographic immunoassay for the qualitative detection of human Myoglobin in whole blood, serum or plasma as an aid in the diagnosis of myocardial infarction (MI).


    Sensitivity and Specificity

    The Myoglobin Rapid Test Device (Whole Blood/Serum/Plasma) has been evaluated with a leading commercial Myoglobin EIA test using clinical specimens. The results show that the sensitivity of the MYO Myoglobin Rapid Test Device (Whole Blood/Serum/Plasma) is 100% and the specificity is 97.8% relative to the leading EIA test.

    Myoglobin Rapid Test vs. EIA

    MethodEIA

    Total

    Results

    Myoglobin Rapid TestResultsPositiveNegative
    Positive59867
    Negative0356356
    Total Results59364423

    Relative Sensitivity: 100% (94.0%-100%)*

    Relative Specificity: 97.8% (95.7%-99.0%)*

    Accuracy: 98.1% (96.3%-99.2%)* 95% Confidence Interval


    Precision

    Intra-Assay

    Within-run precision has been determined by using replicates of 10 tests for each of three lots using Myoglobin specimen levels at 0 ng/mL, 50 ng/mL, 100 ng/mL, 200 ng/mL, and 400 ng/mL. The specimens were correctly identified >99% of the time.

    Inter-Assay

    Between-run precision has been determined by 3 independent assays on the same five specimens: 0ng/mL, 50ng/mL, 100 ng/mL, 200 ng/mL, and 400 ng/mL of Myoglobin. Three different lots of the MYO Myoglobin Rapid Test Device (Whole Blood/Serum/ Plasma) have been tested using these specimens. The specimens were correctly identified >99% of the time.

    Interfering Substances

    The Myoglobin Rapid Test Device (Whole Blood/Serum/Plasma) has been tested and no interference was observed in specimens containing 110 mg/mL human albumin, 6 mg/mL bilirubin, 10 mg/mL hemoglobin, 5 mg/mL cholesterol and 15 mg/mL triglycerides.

    The following compounds have also been tested using the MYO Myoglobin Rapid Test Device (Whole Blood/Serum/Plasma) and no interference was observed.

    AcetaminophenCreatineMetoprolol Tartrate
    Acetoacetic AcidDiclofenacMoracizine Hydrochloride
    Acetylsalicylic acidDigoxinNifedipine
    AnisodamineDL-TyrosineOxalic Acid
    Ascorbic AcidEthanolOxazepam
    AtenololFelodipinePentoxifyline
    Atorvastatin CalciumFlunarizine HydrochloridePhenobarbital
    CaffeineFurosemideQuinine
    CaptoprilGentisic AcidRamipril
    ChloramphanicolHydrochlorothiazideVerapamil
    ChlordiazepoxideIsosorbide Mononitrate
    CilazaprilLabetalol

    Quality Cardiac Myoglobin Uncut Sheet Rapid Test ISO13485 Rtk Blood Test for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: Hangzhou Aichek Medical Technology Co.,Ltd
    *Subject:
    *Message:
    Characters Remaining: (0/3000)
     
    Find Similar Products By Category: